We just bought their Synergy4 and love it. It's a dual monochromator and filter based unit with 7 read outs that meets all of our needs. You can use it with ...
Hi Kevin, I used them a lot in Australia and they are good, nothing complicated, not the top of the line but then you rarely need an expensive instrument to do...
We have used BioTek plate readers for several years now, and I will tell
you that I am very pleased with their performance. We use the BioTek
ELx800 as part...
John Bradshaw
JBradshaw@...
Jun 3, 2008 3:58 am
6080
At our company, we have been struggling to find conditions(pump, pump speed, tubing size, nozzle diameter, etc etc) that allow non contact dispensing of 10 -...
... While I can't give you info on their conditions, several of our customers are using our milliGAT-LF (http://www.globalfia.com/pumps_milligat.html) pump for...
Hi all- Does anyone in the group have information about vendors providing custom Class 100 or 1000 enclosures for 2 stack/4 position Cybio Cybi-well pipettors?...
Hello, In Europe I have used Bigneat who do a good job but I don't have a US contact. Cheers, Neil ... -- -- Neil Benn Msc CEO Ziath Ltd Website -...
Neil Benn
neil.benn@...
Jun 13, 2008 1:30 pm
6084
Hello, I'm not certain, but I think that BioSero may be the US contact. Regards, Kevin McLeaster ... custom ... footprint ... [Non-text portions of this...
Hey Matt & Neil - US distributor for Bigneat biosafety cabinets is BioSero: www.bioseroinc.com <http://www.bioseroinc.com/> East coast contact would be...
Tom Wright, T-US
tom.wright@...
Jun 13, 2008 9:13 pm
6086
Matt, My company can provide such enclosures for your application. You can view our product specific webpage for details,...
we have an assay where we need to aspirate sample containing 0.5 M HCl with steel tips. Is this a problem? Any comments and recommendations are highly...
I have a need to dispense slightly viscous solvent solutions into well plates which is proving somewhat problematic. I have been using a Flexispence HT but get...
Acidic halides corrode steel, even stainless steel. If contamination of your sample with metal ions is not a problem, you may be able to use steel tips on the...
Mr. Abel, I am assuming you mean Stainless Steel SS, but with Stainless?even low M HCL solution may cause a problem. 316 SS is preferable to common 302/303/304...
Yes, prolonged exposure to mineral acids will rot the steel. Brief exposure is OK, for cleaning etc., but when most HPLCs where still steel based it was a big...
even if the tips are ok, the fumes can corrode other things exp aluminium - carefully monitor your equipment ... -- Ezra S Abrams 4 Colbert Rd Newton MA 02465...
ezra s abrams
ezracolbert@...
Jun 17, 2008 1:35 pm
6093
Hello there, I can't find any literature/published data for CV's for a serial dilution. I am using a Tecan and am of the opinion I should be getting better...
Does anyone know of a system that adds sterile media (250ul to 3 ml per well) to a stack of 48,24 or 12 well lidded plates, and prints and applies barcode...
Hello Everyone, Multistep pipetting processes can get ugly fast. Let us assume a best case scenario of 1% per pipetting operation. If your serial dilution...
Hello, has anybody ever used the Matrix Visionmate 96 plus to read bottom 2D barcodes of 24-well racks/plates? I have been told by our local distributor that...
Hi Xavier: It is possible with the newer version of the VisionMate96 Plus software. Give me a shout off-line and I can put you in touch with someone who can...
Hello Mark, My name is Kenny Drew and I am the Automation Sales Manager for ThermoFisher Scientific - Matrix Liquid Handling Equipment. The Matrix product...
The Laboratory Robotics Interest Group Discussion Group is for comments or questions about Laboratory Robotics and Automation. You receive messages posted to...
lrig-discussion@yahoo...
Jul 1, 2008 7:54 am
6100
All, What are the lowest assay volumes successfully used in half area plates, either the ones by Corning or Greiner? The Corning brochure states a range of 75...
Hi JW Not many but a 30 µl assay final volume worked in a Corning half plate for me Aj ... From: janicew3 <janicew3@...> Subject: [lrig-discussion]...
I have successfully used 9 ul final in Greiner. Chakk ________________________________ From: lrig-discussion@yahoogroups.com ...
Ramesha, Chakk
chakk.ramesha@...
Jul 6, 2008 8:46 pm
6103
Hi I'm trying to find a dye that will shift absorbance values when it's temperature changes. In terms of sensitivity, ideally it would be something like...
Hello, As you may or may not be aware, a discussion about Open Source Software has lead to a project startup called Open Lab Automation. The aim of the...
Neil Benn
neil.benn@...
Jul 6, 2008 8:46 pm
6105
Dear colleagues, We have lost a valued partner and friend. John Morin passed away last week. He was a prolific contributor to LRIG, presenting at chapter ...
*I googled your request and immediately found this article. Maybe someone has found a dye with an absorbance profile that shifts...but fluoresence is more...