5475
The Laboratory Robotics Interest Group Discussion Group is for comments or questions about Laboratory Robotics and Automation. You receive messages posted to...
lrig-discussion@yahoo...
Send Email
Jan 1, 2007
12:48 pm
5476
Happy New Year Everyone ! We have a BCI Biomek 2000 and Hereaus (Thermo) Cytomat 6000 we want to integrate to automate feeding mammalian cells. Does anyone...
Tom Strader
ts_8675309
Offline Send Email
Jan 5, 2007
4:11 am
5477
Hello Happy new year and thanks for any help in advance. I was wondering if anyone knew of a bulk dispenser with an integrated stacker that can dispense up to...
D W
biorobotdude
Offline Send Email
Jan 5, 2007
4:13 am
5478
We use the Matrix Wellmate for 1ul dispensing but I haven't used their stacker. The dispenser works great for out needs. Mike Guerin ...
Mike Guerin
mlmguerin
Offline Send Email
Jan 6, 2007
3:58 pm
5479
Hello - We donated our Zymate XP/System V robot to a local school and had it up and running last year. Unfortunately the system no longer works and we are...
mbenn1ng
Offline Send Email
Jan 6, 2007
4:00 pm
5480
DW?, here's another to add to your list. I've never used the stacker however I find the dispenser to perform very well, depending on your liquid. This is...
GRAY (James Gray)
grayj@...
Send Email
Jan 6, 2007
4:01 pm
5481
Hello, BioTek manufactures the mFill Microplate Dispenser for dispensing up to 3 ml into 96-well deep well plates. The mFill is controlled from its onboard...
Jason Greene
GreeneJ@...
Send Email
Jan 6, 2007
4:08 pm
5482
Hi, I'm looking forward to a simple single channel dispenser that can be integrated into an automated system. The ideal characteristics I'm looking for are: ...
iwyates
Offline Send Email
Jan 6, 2007
4:09 pm
5483
Titertek offers two relatively new options. The Multidrop AH, has an adjustable height head that allows pipetting in 96 and 384 wells, in an infinitely ...
Jackbonham@...
Send Email
Jan 6, 2007
7:31 pm
5484
Hi, The Hudson micro10 has many advanced features that outperform syringe and peristaltic pumps. There is never any calibration needed and no valves to fail....
Steven Riedmuller
riedinc
Offline Send Email
Jan 6, 2007
7:34 pm
5485
Check your com ports. Make sure the robot is plugged into the correct port and that the software is trying to access the correct port. da...
david
dr01allen
Offline Send Email
Jan 7, 2007
4:48 am
5486
Hello Here is a single channel dispenser which fulfills some of your requirements. Using the HamiltonCompany MicroLab 500 hardware, it's <$us5000 complete. It ...
John Brohan
jbrohan
Offline Send Email
Jan 8, 2007
5:15 pm
5487
Matrix Wellmate Stackers can drop plates if the retractable pins are stuck due to dried liquid or lint. May or may not be a problem depending on how clean you...
Jason Wu
jay2357913
Offline Send Email
Jan 9, 2007
1:33 am
5488
... It sounds like the Gilson 222XL would fit the bill, but this was discontinued recently so you would probably be looking at its replacement, I think it's...
Tom Hawkins
nekomatic
Offline Send Email
Jan 9, 2007
2:42 pm
5489
Can any vendor that sells low profile, 96 well, filter plates please contact me at the email address below. Thomas R. Todaro Automation Specialist II E:...
Thomas Todaro
ttodaro2001
Offline Send Email
Jan 10, 2007
8:43 pm
5490
I am implementing a plate washer and adapting a manual Elisa protocol to an automated protocol. I am having a problem with high background. The plate comes...
dawnalderman
Offline Send Email
Jan 12, 2007
5:28 am
5491
Hello, I have adapted a manual ELISA protocol to automation. When I found high background issues, it typically came from the HRP-substrate being exposed to...
Dan Rush
dr_loudspeaker
Offline Send Email
Jan 13, 2007
12:47 am
5492
Some Issues I have seen when automating elisas comes from washing the plate higher than the level that is was blocked. If you are using eurpium as your...
D W
biorobotdude
Offline Send Email
Jan 15, 2007
6:46 pm
5493
We have developed a 96-well extraction plate for small molecules that does not require pressure or vacuum systems. The mechanism uses immobilized liquids...
robertwohleb
Offline Send Email
Jan 16, 2007
3:13 am
5494
Try washing the wells 2-3 times with 60 ul (384well) or 150 ul (96well) of 10% "clearcoat" solutions (available from FremontBiosolutions and probably other...
Ramesha, Chakk
chakk.ramesha@...
Send Email
Jan 16, 2007
3:13 am
5495
Hello Rob, Our Zephyr liquid handler (http://www.caliperls.com/products/zephyr_workstation.html) seems to be a pretty good fit for your application. We can...
ndonzel
Online Now Send Email
Jan 17, 2007
2:40 am
5496
Hi, This may be late, but there is a robust solution by Hudson Control Group. Our StackLink stacker actually grips the plates and pulls them out of the stack,...
Steven Riedmuller
riedinc
Offline Send Email
Jan 17, 2007
2:42 am
5497
Hi Rob, At Velocity11 we have the Bravo that would be able to automate your application. http://www.velocity11.com/products/bravo/product_info.html The...
Ian Yates
iwyates
Offline Send Email
Jan 18, 2007
5:50 am
5498
depending on the intensity of the shake this assay could be put on a titertek map c type instrument where the plate is shaken under the wash head. the washer...
D W
biorobotdude
Offline Send Email
Jan 19, 2007
10:55 pm
5499
Hello Rather than re invent the wheel for the nth time I thought I would ask this group for recommendations for Bar coding Master Plates and Daughterplates (if...
Andrew Johnston
hankjam
Offline Send Email
Jan 20, 2007
7:14 pm
5500
... Hi Hankjam we use different pipettors were we reuse tips. We normally pipette DMSO or protein solutions and need to be sure to have no crosstalk, blocked...
atopp2
Offline Send Email
Jan 22, 2007
2:20 pm
5501
Hello, For a relatively small amount of plates - I would simply apply the labels manually. So after neatly sidestepping any questions about applying machines,...
Neil Benn
neil_david_b...
Offline Send Email
Jan 22, 2007
2:27 pm
5502
Hello, i am doing an automated assay using a PW384 plate washer from Tecan that requires large volumes of wash buffer.I run out very quickly and I wanted to...
bei_fengao
Offline Send Email
Jan 22, 2007
8:38 pm
5503
Hi Hankjam I have another possible solution which is our Tipwash ULTRA ultrasonic wash station. I don't want to do too much of a sales pitch, so if you are...
David Ley
david.ley@...
Send Email
Jan 22, 2007
8:39 pm
5504
Hello, Tip washing, as Andreas has described, is time consuming. There is a point of diminishing return when it comes to tip washing. In other words, at some...
Rodriguez, Richard
Richard.Rodriguez@...
Send Email
Jan 22, 2007
8:45 pm